Customization: | Available |
---|---|
Certification: | CE, FDA, ISO13485 |
Type of Body Fluid-Processing Device: | Ingredients Blood Transfusion Device |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
China Hemodialysis Machine Medical Kidney Dialysis Machine Manufacturer
Model : MC-DYM-02
MC-DYM-02 main technical parameter
Volume (length× width ×height):370mm×340mm ×1570mm
Weight: about 90kg
Power supply voltage: AC220V±10%
Frequency: 50HZ ~ 60 HZ
Power: 1500W
Blood pump/spare pump:
Flux: 15 ~ 340ml/min (D 6mm)
20 ~ 460 ml/min (D 8mm)
Heparin pump:
Flux: 0.1ml/h~10ml/h(±5%)
Precise: 0.1ml/h
Injector size: 20ml/30ml/50ml 10ml/20ml/30ml (selectable)
Arterial pressure:
display scope:-300mmHg~+300mmHg (±10mmHg)
Venous pressure:
display scope:-50mmHg~+300mmHg (±10mmHg)
Dialysate flow:300ml/min~800ml/min linearity adjustable(+10%)(-5%)
Temperature:35.0ºC~39.0ºC
Resolving rate:0.1ºC
Conductivity:13mS/cm~15.5mS/cm(±0.1 mS/cm)
UF flow scope: 0~1800ml/h(±30ml/h )
0~4000ml/h(optional)
ISO UF flow scope:0~2000ml/h(±30ml/h)
TMP
Scope:-100mmHg~+600mmHg (±20mmHg)
Blood leakage monitor: over 1ml blood per liter dialysate (flow:500ml/min)
Blood level monitor: ultrasonic sensor
Air bubble monitor:
Infrared and response threshold value:Single air bubble of 200u I exists when Blood flux is 200ml/min
Inflow pressure:0MPa~0.6MPa
Inflow temperature: 5ºC~30ºC
Environment temperature: 10ºC~30ºC, Relative humidity≤70%
Rinse/disinfection: chemical disinfection (citric acid,peracetic acid and oxalic acid)
Hot rinse: 80ºC
Back-up power supply: last for 15-30min after electric-cut.
function
Be applicable of hemodialysis treatment,hemoperfusion,Plasma Exchange,Pure ultrafiltration project.
Touch screen was equipped and it can show Naand UF profiles;In addition,the functions ie.backup power supply,Kt/V,HCO3adjustment,single needle,acetate treatment,pyrogen filter port and net connection ect.are optional.
Clinical Application
Hemodialysis Therapy:same as MC-2000
Pure ultrafiltration Therapy: same as MC-2000
Hemoperfusion Therapy:same as MC-2000
Plasma exchange Therapy:same as MC-2000
Na+and UF profiling:
Be applicable for reducing the cases during dialysis treatment such as low/high pressure,muscle convulsion, headache,nausea,vomit ect.
Kt/V function:
The sufficiency of dialysis can be reflected by Kt/V.
HCO3 adjustment function:
Be applicable for sericous acid toxicosis patient.
Be capable to calibrate the acid toxicosis status according to patient's state of illness.
Backup power-supply:
It can last for 15-30 minutes after power cut off.
Pyrogen filter port:
Completely filtrating the remained bacteria and endotoxin.(optional)
Net connection:
One computer can audit and control the treatment of maximum 20 units hemodialysis machine (optional)
Our hemodialysis machine for kidney in different hospitals worldwide