Customization: | Available |
---|---|
Customized: | Customized |
Certification: | CE, FDA |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Power requirement
|
AC220V-230V/50Hz or AC110-120V, 50/60Hz, 1000VA
|
Control Mode
|
Air mode and Baby mode controlled by micro-computer
|
Control range of air temperature
|
25ºC-37ºC (override mode: 37ºC-39ºC is optional)
|
Control range of baby temperature
|
34ºC-37ºC (override mode: 37ºC-38ºC is optional)
|
Accuracy of skin temperature sensor
|
±0.2ºC
|
Variability of temperature
|
≤0.5ºC
|
Uniformity of temperature
|
≤0.8ºC(mattress is on the horizontal position)
|
Humidity display range
|
0%RH~99%RH
|
Humidity control range
|
20%~90%RH
|
Inclination of the bassinet
|
±12°
|
Noise inside hood
|
≤45dB(A) (Environment noise ≤35dB(A))
|
Failure alarm
|
Over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, system alarm and so on.
|
The average value of total irradiance for bilirubin Ebi max of phototherapy equipment
|
≥10uW/cm2/nm (blue fluorescent lamp)
≥21uW/cm2/nm (blue LED) |
The average value of total irradiance for bilirubin Ebi max of phototherapy equipment under bassinet
|
≥13uW/cm2/nm (with SMD LED)
|
Built-in timer range
|
0~99999.9 hrs
|
Accuracy of weight display
|
±1% (equipped with weighing system)
|