Multi Parameter Signal Patient Simulator Medical ECG Simulator
Model:MCS0398
Product Description
Introduction
It is a portable device, features in small volume and light weight. It can simulate accurately 12-lead ECG, Respiration, Temperature , 2-channel IBP or 4-channel IBP, the default is 4-channel IBP. And it can be tested in where the monitor is used.
Parameters
ECG:
Conventional: All leads have the same voltage with the II-lead as the reference
Amplitude: 0.5, 1.0, 1.5 and 2.0mV
ECG Rate: 15, 20, 25, 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300, 320, 340 and 350 BPM
ECG artifacts: 50 and 60 Hz, muscle, baseline and breathing
The elevation of the ST segment: +0.8 to -0.8 mV, the step is 0.1 mV
Arrhythmia:
ATRIAL: Atrial Fib(coarse), Atrial Fib(fine), Atrial Flutter, Sinus Arrhythmia, Missed Beat, Atrial Tachycardia, Nodal Rhythm, Supraventricular Tachycardia
Ventricular: Vent Fib (coarse), Vent Fib (fine), Premature Ventricular Contraction (PVC) 6/min, PVCs 12/min, PVCs 24/min, Frequent Multifocal, Bigeminy, Trigeminy, Paired PVCs, Run 5 PVCs, Run 11 PVCs, Vent Tachycardia, Asystole
Premature Beat: Atrial Premature Atrial Contraction (PAC), Nodal PAC, Left Ventricular Focus PVC1, Early Left Ventricular Focus PVC1, R on T Left Ventricular Focus, Right Ventricular Focus PVC2, Early Right Ventricular Focus PVC2, R on T Right Ventricular Focus, Multifocal PVC
Conduction Defects: First Degree Heart Block, Second Degree Heart Block, Third Degree Heart Block, Right Bundle Branch Block, Left Bundle Branch Block
Four IBP Channels:
The default: 00/00mmHg
Dynamic IBP(unit: mmHg)
Channel 1: Artery 120/80, Radial 120/80, Left Ventricle 120/00, Right Ventricle 25/00
Channel 2: Artery 120/80, Radial 120/80, Left Ventricle 120/00, Right Atrium(CVP) 15/10, Right Ventricle 25/00, P-ulmonary Artery 25/10, Pulmonary Wedge 10/2, Left Atrium 14/4
Channel 3: Artery 120/80, Radial 120/80, Left Ventricle 120/00, Right Atrium(CVP) 15/10, Right Ventricle 25/00, P-ulmonary Artery 25/10, Pulmonary Wedge 10/2, Left Atrium 14/4
Channel 4:
Auto Swan-Ganz (every 15 seconds): RA(CVP), RV, PA, PAW (Pulmonary Wedge)
Man Swan-Ganz: press ENTER key to change every time
Static-Pressure Levels
Channel 1: 0, 80, 160, 240, 320 and 400mmHg
Channel 2: 0, 50, 100, 150, 200 and 250mmHg
Channel 3: 0, 20, 40, 60, 80 and 100mmHg
Channel 4: P4 tracks the static pressure levels of P3
Respiration:
Respiration Rate: 0 (Apnea), 15, 20, 30, 40, 60, 80, 100, 120 and 150 R/MIN
Impedance: 0.2, 0.5, 1.0 and 3.0 Ω
Baseline impedance: 500, 1000, 1500, 2000Ω
Apnea: 12 s, 22 s, 32 s
Temperature:
Four temperature: 0ºC(32°F), 24ºC( 75°F), 37ºC(98.6°F), 40ºC(104°F)
Simulate the signal of square,triangle, pulse and sine with adjustable frequency.
Simulate the pacemaker signal.
Simulate the signal of standard ECG calibration.
Including the ECG waveforms of ANSI/AAMI EC13.
Accessories |
Standard accessories |
a power adapter(9V/1000mA(CE), 220V input), an User manual of MS400, a plastic bag of 200mm * 150mm, a packaging box, a plastic packaging. |
Optional accessories |
a power adapter(15V/1000mA)(UL), 110V input), a power adapter(15V/1000mA, 220V input) , a temperature cable, an IBP cable, a CD disk(including a case conversion software) |