Item
|
Content
|
Parameters
|
Power supply
|
Voltage
|
380V±38V
|
Frequency
|
50Hz±1Hz
|
|
Capacity
|
≥40kVA
|
|
X-ray high voltage system Radiography
|
Exposure time
|
0.02s-4s, 21 steps
|
Tube voltage
|
44kV-125kV, 41 steps
|
|
Tube current
|
32mA-500mA, adjust stepped
|
|
X-ray high voltage system Fluoroscopy
|
Tube voltage
|
44kV-110kV, 2kV/step
|
Tube current
|
0.5mA-5mA
|
|
X-ray tube
|
Output power
|
Big focus:50Kw small focus:30kW
|
Anode thermal capacity
|
140KJ(180KHU)
|
|
Module thermal capacity
|
1100kJ(1500kHU)
|
|
Rotating anode
|
2800rpm
|
|
Focus
|
2.0mm /1.0mm
|
|
Remote diagnosis table
|
Table rotating range
|
90°~0°~-25°
|
Table surface Horizontal movement
|
±110mm
|
|
Longitudinal movement of frame and spot film device
|
≥720mm
|
|
Film size
|
8″×10 ″-14″×17″
|
|
Spot film division method
|
Whole film, 2-division film, 3-division film, 4-division film
|
|
Control mode
|
Remote control Bed-control Soft boot and stay of frequency conversion
|
|
Limiter
|
electric multi-leaf
|
|
Grid
|
Grid density: 103L/INCH
Grid Ratio :10:1
focusing distance 120cm 15″×18″
|
|
Image system
|
Image intensifier
|
Toshiba 9''
|
CCD camera
|
Imported 0.47 mega pixel
|
|
Monitor
|
14'' Medical high definition monitor, Horizontal centerresolution 1000 line, margin 800 line
video bandwidth12.5MHz, 50bit 25 frame/s
|
|
Digital image processing system(CCU center control)
|
High definition progressive output
8 grades noise reduction
8 images in storage