Customization: | Available |
---|---|
Theory: | Autoclave |
Type: | Pressure Steam Sterilization Equipments |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Parameter |
|
Air removal | Pulse Vacuum (can be set 3-9times) |
Drying | Vacuum drying |
Stand by or not | yes |
The material of the shell | Fully stainless steel |
Adding water automatically | standard |
Exhausting water automatically | standard |
Door lock | Fully automatically, double door interlock |
Outer size(L*W*H) | 1100*1300*1700mm |
Minimum installation room(L*W*H) | 2200*2300*2200mm |
Door Structure | Double door motorized door |
Sterilizing room | 288L±5% |
Sterilizing chamber shape | Horizontal type, 304 type, welding, mechanical polishing,0.4 |
Chamber size (L*W*H) | 800*600*600mm |
Net.Weight | 900kg |
Gross.Weight | 1050kg |
Package size(L*W*H) | 1300*1450*2100mm |
Water tank capacity | 58L |
Average water consuming for each cycle | 40KG |
Maximum temperature of the water tank | 136ºC |
Design pressure | 0.245Mpa |
Design temperature | 139ºC |
Working power supply | AC380V,50HZ, | |||
Power consumption | ≤28KW | |||
Heat average | ≤+2ºC | |||
Steam pressure | 0.3-0.5Mpa | |||
Water resource pressure | 0.1-0.3Mpa | |||
Compressed air | 0.5-0.8Mpa | |||
Power fluctuate | ±10% | |||
Control system | Main controller PLC,microcomputer touch screen,printer and the other control elements |
|||
Display | 7" colorful SIEMENS |
screen, | touchable | from |
With printer | yes | |||
TRAYS (standard allocation) | 2 | |||
Tray size(L*W*H) | 550*760*50mm | |||
Sterilizing cycle | 5 fixed sterilizing cycles+2 fixed testing cycles+1 fixed debug cycle+1 fixed fluid cycle+ free setting cycle |
|||
Sterilizing items | Solid,fluid(heat resistance and placed into un sealed fluid) |
|||
Sterilizing temperature(selectable) | From 105ºC to 134ºC |
|
sting time 5 |
unsealed fluid | ||||||||
BD test | 5 | 3 | 10 | 5 | 134ºC | 3. 5 |
5 | 29.5 | Testing procedure(in accordance with the requirement of the testing bad) |
Vacuum test | Sucking the air till the vacuum rate reach -86Kpa |
In the following 10 mins, the decrease allowable pressure each min is ≤ 0.13KPA, or the total decrease pressure for 10 min is ≤1.3KPA. |
|||||||
Debug cycle | For the use the the factory inspection and repairing | ||||||||
Free cycle | the sterilizing parameter can be set freely in accordance with different requirement . |